Bien Airotor - Prestige L

Bien Airotor - Prestige L

bien-air

Airotor handpiece
New Rs. 57275

Starting at: Rs.57,275.00

EMI starts from Rs.2,777.07   View Plans >

Bien Air Prestige L Air Rotor

Rs.63,930.00

Rs.57,275.00

Bien Air Prestige L Air Rotor + Unifix 4H L

Rs.75,960.00

Rs.68,000.00


OR
Description

Details

  • Type - Turbine with light, small head
  • Connection - Using 4-way Unifix quick-connect coupling, rotating 360 degree
  • Speed of rotation - 335,000 rpm
  • Recommended air pressure - 3.1 bar
  • Cooling - Diffuser, with 3 separate air and water sprays, water antiretraction valve
  • Chuck locking mechanism - Push-button with anti-heating system
  • Light transmission - Multi-strand optical glass conductor
  • Lighting - Halogen
  • Cleaning/Sterilisation - Thermo washer disinfectable / Steam autoclave up to 135 degree C
  • Weight - 73 g
  • Prestige L ref. - 1600380-001
  • Prestige L + 4-way Unifix set ref. - 1700184-001

 

MAINTENANCE OF TURBINES.

Preparation
A.

REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
Tissue_rubbing
B.

PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING
Cleaning_A
A.

CLEAN UNDER RUNNING WATER AT <38°C

Cleaning_B
B.

REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

Cleaning_CSpraynet-Tissue-spray_CTissue_rubbing_C
C.

INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment 1
    (for 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the turbine with a cloth soaked in Spraynet
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
Cleaning_C-primeBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways
C'.

CLEANING THE INSIDE WITH SPRAYNET FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Spraynet attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

DISINFECTION
Disinfectant-Tissue-spray_ATissue_rubbing_A
A.

DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)

Do not immerse the instrument in a disinfectant bath

OR
Thermal-disinfector
B.

THERMAL DISINFECTION (follow the manufacturer's instructions for use)

LUBRICATION
Lubrification_A1Lubrification_A2
A.

LUBRICATION WITH LUBRIMED

OR
Lubrification_B1Support_B
B.

LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before use
  • Use the Lubrifluid attachment 23 ou 4depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
LubrificationBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways
B'.

INTERNAL LUBRICATION WITH LUBRIFLUID FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Lubrifluid attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

STERILISATION
Sterilization
A.

PLACE THE TURBINE IN A STERILISATION BAG

Sterilizer
B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS

  • Instruments bearing the "135°C" symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated

 

 
  • Type - Turbine with light, small head
  • Connection - Using 4-way Unifix quick-connect coupling, rotating 360 degree
  • Speed of rotation - 335,000 rpm
  • Recommended air pressure - 3.1 bar
  • Cooling - Diffuser, with 3 separate air and water sprays, water antiretraction valve
  • Chuck locking mechanism - Push-button with anti-heating system
  • Light transmission - Multi-strand optical glass conductor
  • Lighting - Halogen
  • Cleaning/Sterilisation - Thermo washer disinfectable / Steam autoclave up to 135 degree C
  • Weight - 73 g
  • Prestige L ref. - 1600380-001
  • Prestige L + 4-way Unifix set ref. - 1700184-001

 

MAINTENANCE OF TURBINES.

Preparation
A.

REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
Tissue_rubbing
B.

PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING
Cleaning_A
A.

CLEAN UNDER RUNNING WATER AT <38°C

Cleaning_B
B.

REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

Cleaning_CSpraynet-Tissue-spray_CTissue_rubbing_C
C.

INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment 1
    (for 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the turbine with a cloth soaked in Spraynet
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
Cleaning_C-primeBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways
C'.

CLEANING THE INSIDE WITH SPRAYNET FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Spraynet attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

DISINFECTION
Disinfectant-Tissue-spray_ATissue_rubbing_A
A.

DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)

Do not immerse the instrument in a disinfectant bath

OR
Thermal-disinfector
B.

THERMAL DISINFECTION (follow the manufacturer's instructions for use)

LUBRICATION
Lubrification_A1Lubrification_A2
A.

LUBRICATION WITH LUBRIMED

OR
Lubrification_B1Support_B
B.

LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before use
  • Use the Lubrifluid attachment 23 ou 4depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
LubrificationBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways
B'.

INTERNAL LUBRICATION WITH LUBRIFLUID FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Lubrifluid attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

STERILISATION
Sterilization
A.

PLACE THE TURBINE IN A STERILISATION BAG

Sterilizer
B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS

  • Instruments bearing the "135°C" symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated

 

 
Features
  • QUALITY LED - Adjustable brightness, natural light, no overheating and exceptional resistance to sterilisation
  • DUALOOK - The dual multi-strand optical glass conductor provides the best lighting on the market and is robust enough to withstand all conditions.
  • ACCU-SPRAY - Cooling with 3 sprays with 6 separate air and water nozzles. The ultra-precise nebulisation this offers provides the best possible cooling, and ensures perfect visibility.
  • SEALED HEAD - In combination with the anti-retraction valve, a Tefl on lip prevents any residues or contaminants from entering the instrument.
  • CERAMIC REVOLUTION - Longer service life, greater precision. Less noise, less vibration. Superior quality ceramic ball bearings.
  • ACCU-CHUCK - The most precise bur locking and rotation drive technology in the world. Discover instruments with no imbalance or vibration.
  • COOL TOUCH - Push-button with an anti-heating safety system. This reduces the risk of burning your patients with the head of your instrument. For your comfort and your patients safety.
  • SOFT PUSH Bien- Air's unique push-button insert and clamp locking mechanism. Light pressure is all that's required. Bur changing is quick, comfortable, safe and accurate.
  • FREE MOVE - The Bien-Air turbine-hose combination offers a freedom of movement conducive to working comfort, thanks to the light and flexible rotating coupling.
Key Specifications
  • Authentic
  • Reliable
  • Guarantee impeccable hygiene
  • Simple and Unique
Warranty
Manufacturer/Importer warranty of 12 Months applicable for this product. Service and Warranty are handled by the manufacturer/importer directly for this product
packaging
  • 1 unit
Additional Information

Additional Information

Dispatch Date Next Day
Expiry Date N/A
COD Yes
Reviews

Related Products

There are no products matching the selection.

You may also be interested in

Api Crown And Bridge Preparation Kit (Set Of 14)
api crown and bridge preparation kit (set of 14)
Crown And Bridge Preparation Kit
  • Rs.2,500.00

    Rs.1,785.00

    28.6% off
Gdc Root Canal Plugger Luks -6 (.50mm) (Rcpl2)
gdc root canal plugger luks -6 (.50mm) (rcpl2)
Luks
  • Rs.1,848.00

    Rs.1,070.00

    42.1% off
Shanghai Sheets for Bleaching Trays
shanghai sheets for bleaching trays
Sheets for bleaching tray and night guar...
  • Starting at: Rs.840.00

    23.6% off
Modern Ortho Hyrex RME Screw
modern ortho hyrex rme screw
Hyrex RME Screw
  • Starting at: Rs.1,500.00

    11.8% off
Saeshin Lab Brush Motor (Strong204/102l)
saeshin lab brush motor (strong204/102l)
Lab Brush Motor
  • Rs.6,890.00

    Rs.6,100.00

    11.5% off
Cotisen X-Ray Film Holder
cotisen x-ray film holder
X-Ray Film Holder
  • Rs.100.00

    Rs.55.00

    bulk price: Rs.50.00
    45% off
API Bone Collector
api bone collector
Bone Collector
  • Starting at: Rs.3,360.00

    40% off

Sold Out

Gdc Extraction Forceps Lower Anteriors And Roots - 74 Standard (Fx74s)
gdc extraction forceps lower anteriors and roots -...
Extraction Forceps
  • Rs.1,500.00

    Rs.840.00

    44% off
 EMS AIR FLOW Handy 2 Plus - FT170A
ems air flow handy 2 plus - ft170a
AIR FLOW Handy 2 Plus
  • Rs.44,800.00

Sold Out

Capri Autoclavable Lip & Cheek Retractors Set
capri autoclavable lip & cheek retractors set
Autoclavable Lip & Cheek Retractor
  • Rs.450.00

    Rs.440.00

    bulk price: Rs.430.00
    2.2% off

Sold Out

Recently Viewed Products